Electromagnetic Compatibility at the Mains Ports for Medical Equipments

Authors

  • Mircea Ion Buzdugan Department of Electrical Engineering Technical University from Cluj-Napoca Author
  • Emil Simion Department of Electrical Engineering Technical University from Cluj-Napoca Author
  • Horia Bălan Department of Electrical Engineering Technical University from Cluj-Napoca Author
  • Eugen Baru Datronix Computer Ltd Author

DOI:

https://doi.org/10.24084/repqj07.283

Keywords:

Surge and transient immunity, leakage current, electromagnetic immunity, electromagnetic emissions, Ni Zn ferrites, EMI filters

Abstract

The   paper   presents   the   retro   fitting   EMI   (electromagnetic  interference)  filtering  measures  applied  on  medical  equipments,  in  this  particular  case  on  the  four  cells  galvanic   bath   ATM-31,   produced   by   Datronix   Computer,   Romania,    in    the    attempt    of    achieving    electromagnetic    compatibility compliance for this product.   Due  to  the  specificity  of  medical  equipments,  EMI  mains  filtering  must  meet  more  restrictive  safety  conditions,  as  it  is  presented  in  the  paper,  making  impossible  the  use  of  ready-made typical commercial EMI filters. 

Published

2024-01-24

Issue

Section

Articles